remel TODD HEWITT W. CNA (LIM BROTH), REF R064810; used to isolate Group B Streptococci form clinical specimens containing mixed bacterial flora
Recall
- Recall Number
- Z-2531-2023
- Event Number
- 92880
- Firm
- Remel, Inc
- FEI Number
- 1924669
- Product Code
- JSD
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- August 14, 2023
- Posted
- September 8, 2023
- Address
- 12076 Santa Fe Trail Dr, Lenexa, KS, 66215-3519
Description
remel TODD HEWITT W. CNA (LIM BROTH), REF R064810; used to isolate Group B Streptococci form clinical specimens containing mixed bacterial flora
On lot of Todd Hewitt Broth w/CNA (LIM) may exhibit no recovery of target organism during customer QC testing or use with patient samples
ThermoFisher SCIENTIFIC disseminated a "Medical Device Field Action Recall" to its consignees on 08/14/2023 by US mail. The notice explained the problem, the risk to health, and requested the following actions be taken: - Destroy any remaining affected product in inventory; - A review of reported test results should be evaluated by an appropriate technical expert; - Pass the notice to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred; - Keep the notice on file. The firm will replace the product. If you have any questions, please contact our Technical Services Department at 800-255-6730 (US) or email [email protected].
US Nationwide Distribution
139 units (13,900 tests)