Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036)
Recall
- Recall Number
- Z-2531-2020
- Event Number
- 85799
- Firm
- Philips North America, LLC
- FEI Number
- 1218950
- Product Code
- DSI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 8, 2020
- Terminated
- January 9, 2023
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
Philips M3015A Microstream CO2 Extension Model No. 862393 -used with an lntelliVue X3 Patient Monitor (Model No. 867030) or a lntelliVue MMX Multi-Measurement Module (Model No. 867036)
Affected M3015A extensions are incompatible with the lntelliVue X3 Patient Monitor (Model No. 867030) and lntelliVue MMX multi-measurement server (Model No. 867036).If they are connected to a host monitor, communication with the X3 or MMX could be disrupted, resulting in a loss of monitoring on the host monitor.
Philips issued Customer Information Medical Device Correction (CIL86201894B) via priority mail on 6/8/20. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: To check if this issue affects your M3015A, compare the serial number of your device with the affected serial numbers identified; Use affected M3015A extensions only together with an lntelliVue X2 or MMS.; Do not combine an affected M3015A with an lntelliVue X3 or MMX; Philips Healthcare is distributing with this letter an addendum to the Instructions for Use for patient monitors that include directions for the use of the M3015A (MX100/X3, MX400-800 and MX750/MX850). If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800-722-9377.
Nationwide Foreign: Angola Angola Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana Germany Ghana Gibraltar Greece Greenland Guadeloupe Guatemala Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Lithuania Luxembourg Malaysia Martinique Mexico Morocco Mozambique Namibia Netherlands New Caledonia New Zealand Nicaragua Norway Oman Palestine, State of Paraguay Peru Philippines Poland Portugal Qatar R¿union Russian Federation Saint Pierre and Miquelon Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Turkey Ukraine United Arab Emirates United Kingdom
46,597 Units WW