FDA Recall Terminated

Access Hi-Low Exam Tables, Model 6501-XX

Recall: Z-2531-2018 · Initiated March 29, 2018

Recall

Recall Number
Z-2531-2018
Event Number
80164
Firm
Brewer Company, LLC N88
FEI Number
2182536
Product Code
LGX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 29, 2018
Terminated
August 29, 2019
Address
W13901 Main St, Menomonee Falls, WI, 53051-2330

Description

Access Hi-Low Exam Tables, Model 6501-XX

Reason

One lot, the wire harness was over-crimped on some of the connectors that go to the two vertical side switches on the exam table.

Action

The Brewer Company notified customers, via "MEDICAL DEVICE CORRECTION" letter, on about 03/21/2018. Instructions included to cease distribution of the product, coordinate with Brewer to schedule rework of the cable in affected tables as quickly as possible, and to notify Brewer of customers if the product was further distributed to coordinate rework of those tables. Additionally secondary controls were provided that the operator must press and continually hold the foot control pedal to operate the table, the foot control includes a stop button, depressing any pedal on the foot control will stop the table's automatic movement, and the manual instructs the operator to remove any obstructions from under the exam table prior to movement. Questions can be directed to (888) 273-9371.

Distribution

Devices distributed in WI, TX, MN and IL.

Quantity

7 devices