FDA Recall Open, Classified

Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator

Recall: Z-2530-2023 · Initiated July 26, 2023

Recall

Recall Number
Z-2530-2023
Event Number
92873
Firm
Hamilton Medical, Inc.
FEI Number
2937708
Product Code
CBK
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
July 26, 2023
Posted
September 20, 2023
Address
4655 Aircenter Cir, Reno, NV, 89502-5948

Description

Hamilton-T1, REF: MSP161339/02, ESM Shielding Set, Spare Part for ventilator

Reason

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Action

On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.

Distribution

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Quantity

4 shielding set