FDA Recall Terminated

AB SCIEX QTRAP 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.

Recall: Z-2529-2016 · Initiated June 16, 2016

Recall

Recall Number
Z-2529-2016
Event Number
74520
Firm
Ab Sciex
FEI Number
3008935545
Product Code
DOP
Status
Terminated
Root Cause
Software design
Initiated
June 16, 2016
Posted
August 12, 2016
Terminated
March 2, 2017
Address
500 Old Connecticut Path, Framingham, MA, 01701-4574

Description

AB SCIEX QTRAP 4500MD LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5031231 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.

Reason

MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.

Action

AB SCIEX issued recall notification by letter dated June 16, 2016, to all affected customers. The letter identified the problem, provided temporary measures until a permanent software is available. A response form was to be completed and returned. questions regarding this notice contact SCI EX at +1 289 982 2712.

Distribution

Worldwide Distribution - US (Nationwide) Internationally to Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,

Quantity

28 units