Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Recall
- Recall Number
- Z-2528-2023
- Event Number
- 92873
- Firm
- Hamilton Medical, Inc.
- FEI Number
- 2937708
- Product Code
- CBK
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 26, 2023
- Posted
- September 20, 2023
- Address
- 4655 Aircenter Cir, Reno, NV, 89502-5948
Description
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
Distribution US nationwide, including Puerto Rico. OUS: Not provided
12 units