FDA Recall Open, Classified

LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100

Recall: Z-2527-2024 · Initiated July 2, 2024

Recall

Recall Number
Z-2527-2024
Event Number
94948
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
OWB
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 2, 2024
Posted
August 7, 2024
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

LUMINOS Lotus Max -Intended to visualize anatomical structures by converting an x-ray pattern into a visible image Model Number: 11574100

Reason

The support arm may unintentionally lower resulting in injury to persons when they are under the display ceiling/wall suspension when positioning the displays

Action

Siemens issued Customer Safety Advisory Notice (CSAN) XP017/24/S to Director of Diagnostic Imaging and/ or Radiology Laboratory on 7/2/24. Letter states reason for recall, health risk and action to take: Please visually inspect the mounting screw on the ceiling tube (Figure 1) or wall tube (Figure 2) of the display suspension as far as this is possible from the floor. In case the visual inspection of the mounting screw does not show any abnormalities, you can continue to use the system. Nevertheless, until the system has been checked by Siemens Healthineers customer service with the planned field action, we recommend: - not to swivel the support arm of the display ceiling/wall suspension to the mechanical limit with great force and - to ensure that there is no one under the display ceiling/wall suspension when positioning the displays. This Customer Safety Advisory Notice (XP017/24/S) is being distributed to all potentially affected customers. Siemens Healthineers is preparing a field safety corrective action that will be provided with Update XP018/24/S starting in the third quarter of 2024. The corrective action will involve check and replacement of the attachment of the support arm for the display ceiling/wall suspension by a service technician in an onsite visit. In instances where the mounting screw is broken, the support arm will also be replaced with update XP019/24/S. These field actions will be provided to you free of charge. Once the corrective actions are available, our customer service team will contact you to schedule an appointment to perform the above safety corrective action(s). Once the updates are available and if you would like to make an earlier appointment, please feel free to contact customer service at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification

Distribution

US Nationwide distribution.

Quantity

26 units