FDA Recall Open, Classified

American Contract Systems Biopsy Drape Pack convenience kit

Recall: Z-2523-2024 · Initiated May 6, 2024

Recall

Recall Number
Z-2523-2024
Event Number
94908
Firm
American Contract Systems, Inc.
FEI Number
3003419114
Product Code
FCH
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 6, 2024
Posted
August 7, 2024
Address
2610 Ne Industrial Dr, Ste 220, Kansas City, MO, 64117-2648

Description

American Contract Systems Biopsy Drape Pack convenience kit

Reason

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility

Action

The recalling firm issued recall letters on and dated 5/6/2024 via email or overnight mail. The letter explained the event details, product involved, risk to health, and actions to take. The actions included (1) review your inventory and segregate and quarantine all affected trays. If any of the trays have been further distributed, ensure all end users are notified and maintain records of effectiveness; (2) discard all affected kits and document on the enclosed Response Form; and (3) complete the enclosed Response Form and return it as soon as possible.

Distribution

US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.

Quantity

224