FDA Recall Open, Classified

TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.

Recall: Z-2516-2025 · Initiated July 16, 2025

Recall

Recall Number
Z-2516-2025
Event Number
97367
Firm
PREMIA SPINE LTD Giborey Israel 7 Ramat Poleg Netanya Israel
FEI Number
3012401682
Product Code
QWK
Status
Open, Classified
Root Cause
Process change control
Initiated
July 16, 2025
Posted
September 4, 2025

Description

TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.

Reason

Potential for missing pins at tip of inserter.

Action

On 07/16/2025, Premia Spine Ltd. notified Premia Spine Inc. via phone call. It was requested to quarantine any suspected reworked instruments that are at the warehouse and to initiate a field action and return the affected units of the TOPS Inserter that underwent rework. No customers (hospitals/physicians) were asked to quarantine or return units as all affected units were within Premia Spine's control, however one surgeon was notified, on about 07/31/2025, of the issue and was requested to review the surgical procedure, analyze relevant patient data, including x-ray, and perform additional actions if needed to confirm the pin has not remained in the patient's body.

Distribution

US Nationwide distribution in the state of Connecticut.

Quantity

30 units