IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Recall
- Recall Number
- Z-2510-2024
- Event Number
- 94919
- Firm
- Philips North America Llc
- FEI Number
- 3016618143
- Product Code
- MHX
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- June 28, 2024
- Posted
- August 5, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
IntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
Philips North America issued URGENT Medical Device Correction on 6/25/24 to End-Users. Philips Market Organizations will distribute the letters outside of the U.S. Letter states reason for recall, health risk and action to take: Use only default measurement capabilities, as indicated in Table 1 above, for monitoring purposes. Pass this notice to all those who need to be aware within your organization or to any organization where affected devices have been potentially transferred. A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will reload the current device software to enable missing options (MOS, M06, M20). If you need any further information, please contact your local Philips representative or call 1-800-722-9377.
Nationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China Colombia Denmark France Germany Hong Kong India Indonesia Italy Japan Netherlands Portugal Singapore South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand United Kingdom Vietnam
18 units