FDA Recall Open, Classified

Azure S SR MRI, Model Number W3SR01

Recall: Z-2510-2020 · Initiated June 4, 2020

Recall

Recall Number
Z-2510-2020
Event Number
85816
Firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FEI Number
2182208
Product Code
NVZ
Status
Open, Classified
Root Cause
Software design
Initiated
June 4, 2020
Posted
July 10, 2020
Address
8200 Coral Sea St Ne, Mounds View, MN, 55112-4391

Description

Azure S SR MRI, Model Number W3SR01

Reason

There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management", Atrial Lead Position Check", EffectivCRT" algorithms, and AdaptivCRT"). This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.

Action

The firm initiated their medical device correction on 06/04/2020 by letter. The letter explained the problem and informed the consignee of the available software update (version 3.2.01) along with download instructions.

Distribution

worldwide

Quantity

11953 devices