Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Recall
- Recall Number
- Z-2509-2026
- Event Number
- 99007
- Firm
- Staar Surgical AG Haupt Strasse 104 Nidau Switzerland
- FEI Number
- 3002808259
- Product Code
- MTA
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- April 20, 2026
- Posted
- June 18, 2026
Description
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
On 04/20/2026, the firm sent via FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that STAAR has confirmed that during that period, the printing of the IOD within the Stella Patient View, Order Confirmation, or Shopping Cart, when performed locally (i.e. by doctor/clinic), was under certain circumstances producing an incorrect diagram. The incorrect diagram was produced in certain circumstances when users created multiple Toric lens line items with different axis values on the same patient eye, such as with backup or secondary lens reservations. The root cause has been identified as a software coding error in Stella 2.0 that did not assign a unique identifier to each reservation under a patient s eye. STAAR has modified the software code to assign a unique identifier, thereby preventing this issue when multiple reservations exist. Any IOD printed from Stella 2.0 after 5:30pm PDT on April 16, 2026 would not have been affected by this issue. For Customer's awareness: We are informing you so that a proper clinical decision can be made regarding patient management and follow-up, based on your assessment of the clinical outcome. Please immediately complete and return the Recall Response Form acknowledging receipt of this recall and report any adverse events associated with this event. If a Toric EVO/EVO+ device were implanted using an incorrect IOD printout, there is the potential for the following adverse events: " Refractive surprise " Blurred vision " Glares/Halos " Other visual disturbances Repositioning or removal of the lens may need to be considered in these cases. For Questions - contact STAAR Sales Representative or Customer Service at [email protected] or by phone at 1-800-352-7842 Monday through Friday, 6:00am to 5:00pm PDT.
Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.