Brand Name: MediHoney Product Name: MediHoney Hydrogel Model/Catalog Number: 31620, 31640 Product Description: MEDIHONEY Non-Adhesive Hydrogel Sheet Dressing with Leptospermum Honey is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel gelling agent. The gelling agent is contained within the dressing mass. The dressing contains 100%, Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does require a secondary dressing.
Recall
- Recall Number
- Z-2509-2025
- Event Number
- 97404
- Firm
- Integra LifeSciences Corp. (NeuroSciences)
- FEI Number
- 3003418325
- Product Code
- FRO
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- August 5, 2025
- Posted
- September 3, 2025
- Address
- 1100 Campus Rd, Princeton, NJ, 08540-6650
Description
Brand Name: MediHoney Product Name: MediHoney Hydrogel Model/Catalog Number: 31620, 31640 Product Description: MEDIHONEY Non-Adhesive Hydrogel Sheet Dressing with Leptospermum Honey is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel gelling agent. The gelling agent is contained within the dressing mass. The dressing contains 100%, Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does require a secondary dressing.
Packaging failures were identified which could lead to a breach in the sterile barrier.
On August 5, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions for Customers: If you have units of the impacted product remove them immediately from service and quarantine them. Forward this notice to those who utilize the product so they are aware of this recall and can identify any affected product that may remain in clinical areas. Complete the entirety of the Acknowledgement Form and send via email to [email protected] or FAX to 1-609-750-4220. Keep a copy of the form for your records. Actions for Distributors: If you have products listed in Table 1, remove the product from further distribution. Complete the entirety of the Acknowledgement Form and send via email to [email protected] or FAX to 1-609-750-4220. Keep a copy of the form for your records. Check your customer traceability records and notify them if they have any shipments of above catalog and lot numbers. Modify the acknowledgement form to create one from you to your customers. Collect completed response forms and affected product from your customers and indicate total quantities and lots in distributor reply form. For questions regarding these instructions, please contact Customer Service: Monday - Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873 email: [email protected] Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Integra is dedicated to patient safety and manufacturing excellence. We continue to make numerous quality improvements and investments in our facilities around the world. In addition, we are improving our processes in accordance with applicable regulations. We sincerely apologize for any inconvenience this voluntary recall may cause and thank you for your cooperation in this effort.
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Canada, Australia, New Zealand, Korea, Guam, Malaysia.
5374 units