FDA Recall Open, Classified

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT

Recall: Z-2507-2018 · Initiated May 22, 2018

Recall

Recall Number
Z-2507-2018
Event Number
80423
Firm
COVIDIEN MEDTRONIC
FEI Number
1219930
Product Code
GDW
Status
Open, Classified
Root Cause
Process control
Initiated
May 22, 2018
Posted
July 19, 2018
Address
60 NIDDLETOWN AVE, 2ND F, NORTH HAVEN, CT, 06473

Description

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT

Reason

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Action

On May 22, 2018, all U.S. consignees received an UPDATED URGENT MEDICAL DEVICE RECALL and RETURN FORM notice via Federal Express or certified. The initial notice was sent to impacted customers on April 26, 2018. Customers are advised to take the following required actions: 1. Immediately discontinue use, and quarantine affected products listed 2. Return affected product as indicated. 3. Complete the Recalled Product Return Form even if you do not have inventory and submit by either method below: * Product purchased directly from Medtronic: [email protected] or fax to (800) 895-6140. *Product purchased through distributor: [email protected] or fax it to (203) 492-7719.

Distribution

Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta, Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom