FDA Recall Open, Classified

Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/10, Model Number 722005. 3. Allura Xper FD20, Model Number 722006. 4. Allura Xper FD20 Biplane, Model Number 722008. 5. Allura Xper FD10, Model Number 722010. 6. Allura Xper FD10/10, Model Number 722011. 7. Allura Xper FD20, Model Number 722012. 8. Allura Xper FD20 Biplane, Model Number 722013. 9. Allura Xper FD20 OR Table, Model Number 722023. 10. Allura Xper FD10, Model Number 722026. 11. Allura Xper FD10/10, Model Number 722027. 12. Allura Xper FD20, Model Number 722028. 13. Allura Xper FD20/10, Model Number 722029. 14. Allura Xper FD20 OR Table, Model Number 722035. 15. Allura Xper FD20/20, Model Number 722038. 16. Allura Xper FD20/15, Model Number 722058. 510(k) Numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859.

Recall: Z-2504-2025 · Initiated August 1, 2025

Recall

Recall Number
Z-2504-2025
Event Number
97359
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
August 1, 2025
Posted
September 3, 2025

Description

Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/10, Model Number 722005. 3. Allura Xper FD20, Model Number 722006. 4. Allura Xper FD20 Biplane, Model Number 722008. 5. Allura Xper FD10, Model Number 722010. 6. Allura Xper FD10/10, Model Number 722011. 7. Allura Xper FD20, Model Number 722012. 8. Allura Xper FD20 Biplane, Model Number 722013. 9. Allura Xper FD20 OR Table, Model Number 722023. 10. Allura Xper FD10, Model Number 722026. 11. Allura Xper FD10/10, Model Number 722027. 12. Allura Xper FD20, Model Number 722028. 13. Allura Xper FD20/10, Model Number 722029. 14. Allura Xper FD20 OR Table, Model Number 722035. 15. Allura Xper FD20/20, Model Number 722038. 16. Allura Xper FD20/15, Model Number 722058. 510(k) Numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859.

Reason

Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.

Action

Philips began notifying consignees on about 08/01/2025 via letter. Consignees were instructed to notify all applicable personnel, notify customers if the units have been transferred or distributed, maintain the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system, and to complete and return the response form. Philips will be inspecting affected Nehalem Flexvision-2 PCs and installing the air baffle if it is confirmed that this component is missing. Philips expects to start the implementation of this action in affected devices by Q4 2025. Your local Philips representative will contact you to schedule a visit to perform this activity in your system once available.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AZ, CA, DC, FL, GA, HI, ID, IL, KS, KY, LA, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI, WV and the countries of Argentina, Austria, Australia, Belgium, Switzerland, Chile, China, Colombia, Czechia, Germany, Egypt, Spain, Finland, France, United Kingdom, Hong Kong, India, Italy, Japan, Korea, Republic of, Malaysia, Netherlands, Norway, Oman, Pakistan, Poland, Romania, Saudi Arabia, Sweden, Singapore, Thailand, Vietnam.

Quantity

192 untis