FDA Recall Open, Classified

Computed tomography x-ray systems with software syngo.CT VA20A_SP4a, VA20A_SP5, VA30A_SP2, VA30A_SP2a, VA30A_SP3,VA30A_FP2 in Somatom systems: SOMATOM go.Up - Model 11061620 SOMATOM go.Up - Model 11061628 SOMATOM go.All - Model 11061630 SOMATOM go.Top - Model 11061640 SOMATOM X.cite - Model 11330001 Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Recall: Z-2500-2021 · Initiated August 18, 2021

Recall

Recall Number
Z-2500-2021
Event Number
88581
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAK
Status
Open, Classified
Root Cause
Software design
Initiated
August 18, 2021
Address
40 Liberty Blvd, Malvern, PA, 19355-1418

Description

Computed tomography x-ray systems with software syngo.CT VA20A_SP4a, VA20A_SP5, VA30A_SP2, VA30A_SP2a, VA30A_SP3,VA30A_FP2 in Somatom systems: SOMATOM go.Up - Model 11061620 SOMATOM go.Up - Model 11061628 SOMATOM go.All - Model 11061630 SOMATOM go.Top - Model 11061640 SOMATOM X.cite - Model 11330001 Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Reason

Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows, resulting in delay in diagnosis or scan aborts with the necessity for patient rescan may occur

Action

Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice (CSAN) to US customers via CT024/21/S on 8/18/2021. The letter states reason for recall, helath risk and action to take: There are no workarounds available for the identified issues. Siemens Healthineers has developed software update syngo.CT VA30A_SP4 to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also provide workflow improvements, bug fixes for the performance and stability problems observed in the installed base. The corrective action will be provided free of charge and will be distributed via of the update packages described above, depending on your current software version. Following the corrective action, the cause has been eliminated and recurrence of the identified issues are prevented. This software update will be provided to you free of charge. Software updates CT022/21/S and CT025/21/S will be performed onsite. Updates CT016/21/S and CT018/21/S will be performed remotely. The software updates process will require approximately 180 minutes for completion. Please make sure the system and power are stable before and during the process. Please do not switch off the system during the update process. Siemens highly recommends starting the installation when the scanner is not in use or when the necessary time for the update to be completed can be scheduled.

Distribution

US Nationwide distribution.

Quantity

31 units US