LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is indicated for use on patients 1 year of age or older in cardiopulmonary arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). The cprCOACHTM Feedback Technology in the LIFEPAK CR2 defibrillator is indicated for use on cardiopulmonary arrest patients and provides CPR guidance in accordance with the AHA Guidelines for patients 1 year of age or older. The LIFEPAK CR2 defibrillator is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support/AED, advanced life support, or a physician-authorized emergency medical response training program.

Recall: Z-2499-2025 · Initiated July 22, 2025

Recall

Recall Number
Z-2499-2025
Event Number
96949
Firm
Physio-Control, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 22, 2025
Posted
September 2, 2025
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is indicated for use on patients 1 year of age or older in cardiopulmonary arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). The cprCOACHTM Feedback Technology in the LIFEPAK CR2 defibrillator is indicated for use on cardiopulmonary arrest patients and provides CPR guidance in accordance with the AHA Guidelines for patients 1 year of age or older. The LIFEPAK CR2 defibrillator is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support/AED, advanced life support, or a physician-authorized emergency medical response training program.

Reason

Due to required inspections not being performed on products/units that have gone through servicing.

Action

On 07/22/2025, the firm sent via certified mail an "URGENT: Medical device recall" Letter to customers informing them that, Stryker discovered a subset of work orders performed between August 10th 2022 and October 30th 2023 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work order to ensure the devices are performing as intended. Customers are instructed to: 1. Immediately check their inventory to locate the affected product as listed in the letter. 2. On the business reply form, customer need to indicate if their devices have been serviced since October 30th 2023 (after that time, if affected device have undergone Preventative Maintenance through Stryker, Stryker has verified that the PIPs were completed and passed in accordance with the required standards. For these devices No additional action is required. 3. Return business reply form. Upon receipt, Stryker will contact customers to arrange for corrective action on the device if service has not been performed. 4. There are no additional actions that users can and should take. Continue to follow the Operating Instructions and perform the daily checks and user tests. Tests can be found in the chapter "Maintaining the Equipment' subsection "General Maintenance and Testing" For Questions or concerns, contact Customer Service at +1 800 787-9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time, Monday thru Friday or email [email protected]

Distribution

U.S. Nationwide distribution in the states of CO and MT.

Quantity

1 unit