FDA Recall Terminated

cobas SARS-CoV-2 & Influenza A/B Test for use on the cobas Liat System, Catalog Number 09211101190

Recall: Z-2498-2021 · Initiated August 4, 2021

Recall

Recall Number
Z-2498-2021
Event Number
88600
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
QLT
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
August 4, 2021
Terminated
February 23, 2024
Address
1080 US Highway 202 S, Branchburg, NJ, 08876-3733

Description

cobas SARS-CoV-2 & Influenza A/B Test for use on the cobas Liat System, Catalog Number 09211101190

Reason

Customers have reported an increased number of false positive SARS-CoV-2 results.

Action

On August 4, 2021, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were instructed to take the following actions: - Immediately discontinue the use of the cobas SARS-CoV-2 & Influenza A/B Test for use on the cobas Liat system, lot number 10119Z. - For handling previous positive results generated with the cobas SARS-CoV-2 & Influenza A/B test for use on the cobas Liat system, follow your laboratory s standard operating procedures to investigate the potential for false positive results. Please consult with the physician or clinicians at your facility to determine specific clinical implications for your patients. - Repeat testing can be performed using other cobas SARS-CoV-2 & Influenza A/B lot numbers. - Complete all sections of the enclosed faxback form and fax it to 1-888-312-5960 or email it to [email protected]. - Discard lot number 10119Z according to your local waste management ordinance. - File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-800-5973 if you have questions about this recall.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, CA, IL, IN, KS, ME, MI, MN, MO, NY, OR, TN, TX, WA, and WI. The countries of Canada, UK, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Netherlands, Norway, Poland, Slovakia, and Spain.

Quantity

211 kits (US); 3073 kits (OUS)