FDA Recall Terminated

Lyophilized Mouse Monoclonal Antibody Thyroid Transcription Factor-1; 0.1 mL and 1 mL lyophilized tissue culture supernatant in vials; an in vitro diagnostic reagent for the qualitative identification by light microscopy of thyroid transcription factor-1 molecules in paraffin-embedded tissue by immunohistochemical staining; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom; Catalog No.: NCL-TTF-1

Recall: Z-2498-2010 · Initiated August 2, 2010

Recall

Recall Number
Z-2498-2010
Event Number
56406
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
NJT
Status
Terminated
Root Cause
Device Design
Initiated
August 2, 2010
Posted
September 25, 2010
Terminated
January 29, 2012
Address
2345 Waukegan Rd, Bannockburn, IL, 60015-1515

Description

Lyophilized Mouse Monoclonal Antibody Thyroid Transcription Factor-1; 0.1 mL and 1 mL lyophilized tissue culture supernatant in vials; an in vitro diagnostic reagent for the qualitative identification by light microscopy of thyroid transcription factor-1 molecules in paraffin-embedded tissue by immunohistochemical staining; Leica Biosystems Newcastle Ltd., Balliol Business Park West, Benton Lane, Newcastle Upon Tyne NE12 8EW, United Kingdom; Catalog No.: NCL-TTF-1

Reason

In combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor types other than is currently indicated on the instructions for use.

Action

Leica Microsystems sent Medical Device Field Correction letters dated 7/28/10 to the affected customers via first class mail on the same date, informing them that in combination with a sensitive polymer detection system, the TTF-1 antibodies may stain additional tumor types other than is currently indicated on the Instructions for Use. Currently they state that lung and thyroid tumors may stain, but Leicas investigation further characterized the expression of the TTF-1 antigen and it may also be present in other tumors, e.g. thymomas. Customers who have received the affected lot numbers of the three products were requested to refer to the new Instructions for Use to review the amendments made to the Results Expected via the following links: http://bsd.leica-microsystems.com/pdfs/products/ttf-1-ce.pdf; http://bsd.leica-microsystems.com/pdfs/products/ttf-1-l-ce.pdf; and http://bsd.leica-microsystems.com/pdfs/products/pa0364.pdf; The accounts were asked to complete the attached Acknowledgement Form, acknowledging the receipt and understanding of the letter, listing the lot numbers of the units at their facility and fax it back to Leica at 1-847-607-3147. Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.

Distribution

Nationwide and Canada

Quantity

860 units Total for all three products