FDA Recall Open, Classified

BD PYXIS MEDBANK TWR MN CR-2HH-6FH, REF: 169-88 containing software version 3.9.1.9

Recall: Z-2493-2024 · Initiated June 24, 2024

Recall

Recall Number
Z-2493-2024
Event Number
94947
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
BRY
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
June 24, 2024
Posted
August 2, 2024
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD PYXIS MEDBANK TWR MN CR-2HH-6FH, REF: 169-88 containing software version 3.9.1.9

Reason

Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.

Action

On July 11, 2024, CareFusion, subsidiary of Becton, Dickinson and Company (BD) issued a "Urgent Medical Device Correction" Notification, via mail. On June 24, 2024, BD issued Customer Release Notes via their firm's customer portal. BD ask consignees to take the following actions: Clinicians: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. Users should verify the contents of the bin location being restocked. 2. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Pyxis" MedBank" CUBIE" Replenishment Station devices to. 3. Please promptly complete and return the enclosed Customer Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter Distributor: 1. Please provide a copy of the attached customer letter to all facilities within your distribution network that have any product listed in Appendix 1. 2. Please promptly complete and return the enclosed Distributor Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Quantity

1 unit