Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
Recall
- Recall Number
- Z-2489-2023
- Event Number
- 92772
- Firm
- Accuray Incorporated
- FEI Number
- 3003873069
- Product Code
- IYE
- Status
- Completed
- Root Cause
- Device Design
- Initiated
- August 3, 2023
- Posted
- August 31, 2023
- Address
- 1240 Deming Way, Madison, WI, 53717-1954
Description
Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.
It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.
The recalling firm began issuing letters dated 7/19/2023 via hand-delivery by the field service engineer. The letter explained the issue and explained that if their Radixact System is being serviced by a third-party provider, please ensure they are also aware and take relevant precautions while servicing the system. The product impacted by the issue was detailed. The letter explained the issue arises only during the servicing of the Radixact System and customers can continue the clinical use of their system. The letter informs the consignee that Accuray will inspect the upper roller bracket of each impacted Radixact system to confirm there is adequate clearance for a jib crane. If there is sufficient clearance, then no action is necessary. If there is insufficient clearance, then the old bracket will be replaced with the new bracket following the released updated service procedures. An acknowledgment form was enclosed documenting the consignee confirms understanding the content of the field safety notice and has distributed the information to all applicable members of their staff.
Worldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ, OR, PA, SC, SD, TX, and WA. There was government distribution and no military distribution. The countries of Australia, Belgium, Canada, Colombia, France, India, Indonesia, Italy, Japan, Kazakhstan, Malaysia, Poland, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.