FDA Recall Terminated

HeartStart MRx Monitor/Defibrillator Model # M3535A - Product Usage: The HeartStart MRx is a lightweight, portable external defibrillator, offering two modes of operation for defibrillation: manual mode and semi-automatic mode (AED). In manual mode, the HeartStart MRx is a full featured manual defibrillator, designed for use by clinicians trained in Advanced Cardiac Life Support (ACLS). Manual operation allows users to select energy levels for external and internal defibrillation. Users may also perform synchronized cardioversion. Non-invasive external pacing is a device option. In AED mode, the HeartStart MRx allows the provider who is trained in Basic Life Support (BLS) to provide defibrillation therapy. The device analyzes a patient s rhythm and advises the user to deliver a shock. Voice prompts guide the user through the defibrillation process by providing instructions and patient information. The voice prompts are reinforced by messages that appear on the display. In both modes of operation, the HeartStart MRx utilizes impedance compensating biphasic truncated exponential therapy waveform. The HeartStart MRx can also be used for ECG monitoring of a patient using 3, 5 or 12 lead cables.

Recall: Z-2488-2019 · Initiated July 16, 2019

Recall

Recall Number
Z-2488-2019
Event Number
83576
Firm
Philips North America, LLC
FEI Number
1218950
Product Code
MKJ
Status
Terminated
Root Cause
Other
Initiated
July 16, 2019
Terminated
October 20, 2022
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

HeartStart MRx Monitor/Defibrillator Model # M3535A - Product Usage: The HeartStart MRx is a lightweight, portable external defibrillator, offering two modes of operation for defibrillation: manual mode and semi-automatic mode (AED). In manual mode, the HeartStart MRx is a full featured manual defibrillator, designed for use by clinicians trained in Advanced Cardiac Life Support (ACLS). Manual operation allows users to select energy levels for external and internal defibrillation. Users may also perform synchronized cardioversion. Non-invasive external pacing is a device option. In AED mode, the HeartStart MRx allows the provider who is trained in Basic Life Support (BLS) to provide defibrillation therapy. The device analyzes a patient s rhythm and advises the user to deliver a shock. Voice prompts guide the user through the defibrillation process by providing instructions and patient information. The voice prompts are reinforced by messages that appear on the display. In both modes of operation, the HeartStart MRx utilizes impedance compensating biphasic truncated exponential therapy waveform. The HeartStart MRx can also be used for ECG monitoring of a patient using 3, 5 or 12 lead cables.

Reason

Non-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices.

Action

1. Locate the MRx devices M3535A with the serial numbers US00539461 and US00539462. 2. Remove them from clinical use and quarantine these two devices in a secure location. 3. A Philips representative will contact you to arrange for the removal of the affected HeartStart MRx devices and return them to the factory. If you need further information or support concerning this notification, please contact your local Philips representative or call us at 1-800-722-9377.

Distribution

US Nationwide distribution in the state of MA.

Quantity

2 units