MEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A
Recall
- Recall Number
- Z-2485-2026
- Event Number
- 98994
- FEI Number
- 1417592
- Product Code
- OEQ
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 4, 2026
- Posted
- June 17, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
MEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
72 units