Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.

Recall: Z-2485-2012 · Initiated August 23, 2012

Recall

Recall Number
Z-2485-2012
Event Number
63073
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
HXX
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
August 23, 2012
Posted
September 28, 2012
Terminated
August 27, 2013
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.

Reason

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Action

Biomet sent an "URGENT MEDICAL DEVICE RECALL NOTICE'' dated August 23, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. A Fax Back Response Form was included for customers to complete and return via fax to 574-372-1683. Contact the firm at 574-372-1570 for questions related to this recall.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AL, GA, MT, TA, AZ, WA, CA, TA, IN, NY, NJ, VA, MA, MI, MD, PA, KY, NC, UT, NH, FL, OK, CO, LA, OH, MO and SD, and the countries of Japan, Netherlands, Australia, and Costa Rica.

Quantity

268 units total