HS Hi-R NEO 900 Reference # 657 820 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.
Recall
- Recall Number
- Z-2482-2019
- Event Number
- 83571
- Firm
- Haag-Streit USA Inc
- FEI Number
- 1000136533
- Product Code
- EPT
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- July 31, 2019
- Terminated
- April 16, 2020
- Address
- 3535 Kings Mills Rd, Mason, OH, 45040-2303
Description
HS Hi-R NEO 900 Reference # 657 820 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.
software error in the central control unit of the floor stand.
1. Do not perform any other functions at the same time when operation the XY control, such as changes in the light intensity. 2. Perform a software update of the floor stand. you will receive the software version 11.4 through your local representative which will fix the bug. 3. please send us the enclosed response form filled In as soon as possible. 4. If you have passed the products to third parties, please forward a copy of this information and inform the below mentioned contact person. The software update will be organized through Haag-Streit USA. For questions concerning this procedure, please contact: Brian Gallagher Phone: (513) 486-1522 Email: fieldsafetynotice@haag-streit-usa .com
US in the state of OH
0