FDA Recall Open, Classified

BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower / 10885403512674 / 352 BD Pyxis" Anesthesia Station ES / 10885403477836 / 327 BD Pyxis" MedStation" ES (Main) / 10885403512667 / 323 BD Pyxis" CII Safe ES / 10885403512605 / 1116-00 BD Pyxis" Enterprise Server / 10885403518348 and 10885403520341 / 1115-00 BD Pyxis" CIISafe, V9.X, SGL IM, BIO, / 10885403512520 / 111-221 BD Pyxis" CIISafe, V7.X, DBL INTG MAIN, BIO, SOLID DO / N/A / 111-194 BD Pyxis" MedBank Tower / 10885403512568 / 169-29 BD Pyxis" CIISafe, V8.X Desktop PC, Bio / N/A / 107-245-01 Supply Server / N/A / N/A CCE Enterprise SW Site License (2-5) / N/A / N/A BD Pyxis" Logistics / N/A / 806 BD Pyxis" SupplyStation" V9 / N/A / 317 Security Module / N/A / 139698-01 Supply itParActive RF V11 / N/A / 138566-01 BD Knowledge Portal for Medication Technologies / N/A / 136607-01 STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-02 PRS STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-01 STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-02 PRS STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-01 BD Pyxis" CII Safe ES, Desktop PC / N/A / 1156-00 BD Pyxis" SupplyStation" Panel PC V11 / N/A / 1132-00 Pyxis" SupplyRoller V11 / N/A / 1127-00 Dell 640 Server / N/A / 1128-00 BD Pyxis" SupplyStation" System V11 / N/A / 1123-00 BD Pyxis" SupplyRoller V10 / N/A / 1101-00 BD Pyxis" SupplyStation" System V10 / **** / 347 BD Pyxis CII Safe,V9.X DESKTOP PC / **** / 107-252-01 BD Pyxis CII Safe,V7.X,DBL INTG MAIN,BIO,CLEAR DO 111-191 BD Pyxis CII Safe, V9.X, DBL IM, BIO, SLD DOORS / **** / 111-215 BD Care Coordination Engine (CCE) / **** / 134056-01 The Pyxis devices are automated medication dispensing cabinet that is intended to securely store and dispense medications to a qualified and authorized healthcare provider.

Recall: Z-2477-2025 · Initiated May 8, 2025

Recall

Recall Number
Z-2477-2025
Event Number
96477
Firm
CareFusion 303, Inc.
FEI Number
2016493
Product Code
BRY
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
May 8, 2025
Posted
August 28, 2025
Address
10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386

Description

BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower / 10885403512674 / 352 BD Pyxis" Anesthesia Station ES / 10885403477836 / 327 BD Pyxis" MedStation" ES (Main) / 10885403512667 / 323 BD Pyxis" CII Safe ES / 10885403512605 / 1116-00 BD Pyxis" Enterprise Server / 10885403518348 and 10885403520341 / 1115-00 BD Pyxis" CIISafe, V9.X, SGL IM, BIO, / 10885403512520 / 111-221 BD Pyxis" CIISafe, V7.X, DBL INTG MAIN, BIO, SOLID DO / N/A / 111-194 BD Pyxis" MedBank Tower / 10885403512568 / 169-29 BD Pyxis" CIISafe, V8.X Desktop PC, Bio / N/A / 107-245-01 Supply Server / N/A / N/A CCE Enterprise SW Site License (2-5) / N/A / N/A BD Pyxis" Logistics / N/A / 806 BD Pyxis" SupplyStation" V9 / N/A / 317 Security Module / N/A / 139698-01 Supply itParActive RF V11 / N/A / 138566-01 BD Knowledge Portal for Medication Technologies / N/A / 136607-01 STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-02 PRS STOCKSTN V10.X WRD/WL SCANNERS / N/A / 132-53-01 STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-02 PRS STOCKSTN V10.X WIRED SCANNER / N/A / 132-52-01 BD Pyxis" CII Safe ES, Desktop PC / N/A / 1156-00 BD Pyxis" SupplyStation" Panel PC V11 / N/A / 1132-00 Pyxis" SupplyRoller V11 / N/A / 1127-00 Dell 640 Server / N/A / 1128-00 BD Pyxis" SupplyStation" System V11 / N/A / 1123-00 BD Pyxis" SupplyRoller V10 / N/A / 1101-00 BD Pyxis" SupplyStation" System V10 / **** / 347 BD Pyxis CII Safe,V9.X DESKTOP PC / **** / 107-252-01 BD Pyxis CII Safe,V7.X,DBL INTG MAIN,BIO,CLEAR DO 111-191 BD Pyxis CII Safe, V9.X, DBL IM, BIO, SLD DOORS / **** / 111-215 BD Care Coordination Engine (CCE) / **** / 134056-01 The Pyxis devices are automated medication dispensing cabinet that is intended to securely store and dispense medications to a qualified and authorized healthcare provider.

Reason

Antivirus software was not consistently installed on impacted devices during the implementation process.

Action

On 05/08/2025, the firm sent via mail and email an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter informing customers of two issues. For this recall the issue pertains to technicians failing to install (ESET) antivirus software on affected devices. This can potentially result in a malicious actor that could exploit a cybersecurity weakness on a device that is configured without antivirus protection within a hospital network. Customer are instructed and/or informed: For Clinical Users: -to utilize user facility policies and procedures for cybersecurity issues or disaster recovery. -to remove devices from network and continue to use operational devices in critical override per the BD Pyxis" MedStation" ES user guide DME 10000430667, version 00 (pages 74-75). For IT, Safety and Security: -Firewall Protection - The hospital firewall can block vulnerable ports. -Limit physical access to only authorized personnel. -Tightly control management of system passwords provided to authorized users. -Monitor and log network traffic attempting to reach the affected products for suspicious activity. -Isolate affected products in a secure VLAN or behind firewalls with restricted access that only permits communication with trusted hosts in other networks when needed. -Physical security - Manage who can access network or server rooms to reduce cyberattack or data breach risks. -Secure user access implement secure user passwords and/or biometric logins at the application level. -Regular database backups (ES devices only). -Additional Security layers - There are other layers of security still in place to mitigate potential malicious activity on ES devices such as secure Operating System passwords, database and file encryption, and secure user passwords / biometric logins at the application level. For questions: BD Remediation Support Phone: 1-866-583-8783 Phone Hours: 5:00am PST to 4:00pm PST (8:00am ET to 7:00pm ET) Monday - Friday Recall Customer Response Forms - email: BDRC22

Distribution

Worldwide - U.S Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of United Arab Emirates, Argentina, Australia, Belgium, Bahrain, Bermuda, Brazil, Bahamas, Canada, Switzerland, Germany, Egypt, Finland, France, United Kingdom, Hong Kong, Indonesia, Ireland, India, Italy, Jordan, Lebanon, Morocco, Mexico, Malaysia, New Zealand, Qatar, Saudi Arabia, Singapore, Thailand, Greece, Oman, and Taiwan.

Quantity

8353 devices (6738 U.S. and 1615 O.U.S.)