The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Recall
- Recall Number
- Z-2476-2021
- Event Number
- 88546
- Firm
- Philips North America Llc
- FEI Number
- 3016618143
- Product Code
- OWB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 13, 2021
- Terminated
- March 11, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: "The Azurion series can be used in a hybrid Operation Room. "The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Philips notified accounts by Urgent Medical Device Recall letter via certified mailing United States Postal Service (USPS) on 8/13/21. Letter states reason for recall, health risk and action to take: Please place this Medical Device Recall letter with the documentation of the system until this correction has been implemented in your system. Additionally, please return the attached reply form to Phhilps to confirm that the users of the system have reviewed and understood this Medical Device Recall. Philips will correct the affected systems by replacing the ADUs. A Philips service representative will contact you to schedule the correct ion of your system. Please be assured that maintaining a high level of safety and quality is our highest priority. lf you need any further information or support concerning thls issue, please contact your local Philips representative: 1-800-722-9377 (reference FCO72200483
US Nationwide distribution in the states of ID, MA, MI, TX, VA.
1 unit