FDA Recall Terminated

cobas e 601 module (cobas 6000 Modular Series system) Part Number: 04745922001

Recall: Z-2476-2019 · Initiated July 18, 2019

Recall

Recall Number
Z-2476-2019
Event Number
83525
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
JJE
Status
Terminated
Root Cause
Device Design
Initiated
July 18, 2019
Terminated
June 15, 2020
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

cobas e 601 module (cobas 6000 Modular Series system) Part Number: 04745922001

Reason

Quality issue with high pressure solenoid valves may cause inaccurate results.

Action

1. For the cobas 8000 ISE, cobas c 701, and cobas c 702 modules and the cobas c 513 analyzer, stop using these units immediately. Your FES will be in contact with you to replace these affected high pressure solenoid valves immediately. 2. In the event that the high pressure solenoid valves cannot be replaced immediately, for the cobas e 601 module, follow the wash station checks outlined in this Urgent Medical Device Correction (UMDC). Perform these checks once per shift. If these checks show favorable outcomes and quality control results are within range, patient results may be reported. If these checks produce unfavorable outcomes, stop using your cobas e 601 module and call the Roche Support Network Customer Support Center at 1-800-428-2336. 3. Complete the attached fax form and fax or email it according to the instructions on the form. 4. File this Urgent Medical Device Correction (UMDC) for future reference Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.

Distribution

State NY MT HI IN MD TX OH KS KY CA FL OR LA MO AK

Quantity

1 unit