00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry y and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Recall
- Recall Number
- Z-2474-2012
- Event Number
- 62110
- Firm
- Zimmer, Inc.
- FEI Number
- 1000220733
- Product Code
- HWT
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 1, 2011
- Posted
- September 28, 2012
- Terminated
- May 16, 2014
- Address
- 345 E Main St, Warsaw, IN, 46580-2746
Description
00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry y and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Zimmer, Inc. sent an E-mail letter dated April 21, 2011, to all affected customers in the US. A June 2011 E-mail was sent to all customers outside the US. The letters requested Inventory & Operations Managers to check product for proper fit /function, complete the return form with worn provisional heads to Zimmer Inc, Attention Product Service, 1777 West Center St. Warsaw, IN 46580. For questions regarding this recall call 574-372-4753.
Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
231, 641 all lots