Philips Integris XPER FD10 F/C Angiographic X-ray system. This system with Velara Generator can be identified as type number 9890 000 6501 * (number located on the generator cabinet). Intended use for Angiographic X-ray system with and without a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.
Recall
- Recall Number
- Z-2474-2010
- Event Number
- 50114
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 23, 2008
- Posted
- September 22, 2010
- Terminated
- August 10, 2011
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
Philips Integris XPER FD10 F/C Angiographic X-ray system. This system with Velara Generator can be identified as type number 9890 000 6501 * (number located on the generator cabinet). Intended use for Angiographic X-ray system with and without a solid state X-ray imaging device for cardiovascular and vascular diagnostic and interventional procedures.
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
The firm, Philips, sent "URGENT-Device Correction" letters dated November 10, 2008, to their customers and informed them about the reliability issue in the Allura XPER and Integris X-ray systems with the High Voltage Velera (HV) Generator. This reliability issue can lead to a system error causing intermittent loss of fluro and/or exposure acquisition, potentially resulting in the loss of the X-ray image on the display. The customers are advised if they experience High Voltage Generator reliability issue and the HV Generator stops working generating an O3EA error, a cold restart is required and reboot the system when the X-Ray image disappears. The firm will implement multiple mandatory Field Change Orders (FCOs) to solve the Generator reliability by updating the software for each X-ray system. The customers could contact Philips Care Center at 800-722-9377, #5, #4, #1 for any questions regarding this recall or 425-487-7469.
Worldwide distribution.
206 units in US