FDA Recall Open, Classified

Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code: LDLC Siemens Material Number (SMN): 11537214

Recall: Z-2472-2023 · Initiated July 10, 2023

Recall

Recall Number
Z-2472-2023
Event Number
92800
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JFM
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
July 10, 2023
Posted
August 28, 2023
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code: LDLC Siemens Material Number (SMN): 11537214

Reason

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica CH Total Bilirubin_2 (TBil_2), Atellica CH LDL Cholesterol (LDLC), Atellica CH Gamma-Glutamyl Transferase (GGT) and Atellica CH HDL Cholesterol (HDLC)

Action

Siemens Healthineers issued Urgent Medical Device Correction ACHC23-01.C.US to US affected customers via FedEx beginning on 07/10/2023; and Urgent Field Safety Notice ACHC23-01.C.OUS,OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 07/10/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Perform the instructions provided in the Additional Information section. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: " Perform testing of TBil_2 on a separate analyzer from Glucose Oxidase (GluO) " Perform testing of LDLC on a separate analyzer from Lipase (Lip) " Perform testing of GGT on a separate analyzer from Magnesium (Mg) " Perform testing of HDLC on a separate analyzer from Uric Acid (UA) If you choose not to or are unable to separate the assays as indicated above, batch testing of GluO, Lip, Mg and UA may be performed. Note: When batch testing impacted assays, GluO, Lip, Mg and UA, an RPC2 wash mitigation must be initiated after completion of tests with the assay causing carryover, TBil_2, LDLC, GGT and HDLC, respectivel

Distribution

US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam

Quantity

6733 units