FDA Recall Open, Classified

ConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS5-120LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Recall: Z-2470-2026 · Initiated May 14, 2026

Recall

Recall Number
Z-2470-2026
Event Number
98854
Firm
ConMed Corporation
FEI Number
3001451571
Product Code
HIF
Status
Open, Classified
Root Cause
Process change control
Initiated
May 14, 2026
Posted
June 17, 2026
Address
525 French Rd, Utica, NY, 13502-5945

Description

ConMed AirSeal 5mm Access Port and Low Profile Obturator with Bladeless Optical Tip, 120mm Length. Catalog Number: iAS5-120LP. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

Reason

Potential for overpressure alerts.

Action

ConMed notified consignees on about 05/14/2026 via emailed and mailed letter titled URGENT: Medical Device Correction. Consignees were made aware of the issue and risk associated with the issue. They were instructed that a replacement AirSeal disposable should be available prior to procedures in order to minimize potential delays. Before initiating insufflation, the user should ensure the tubing is properly connected between the AirSeal tube set and Airseal Access Ports. If the connection cannot be securely tightened and the overpressure alert cannot be resolved, please replace the affected device. Detailed instructions were provided on how to connect the AirSeal tube set and Airseal Access Ports. Consignees were also instructed to review inventory for affected units, place a copy of the provided Instructions for Use, Part No. AS-DIS-IFU, Rev. B with any inventory in possession with an active expiration date, contact all applicable departments and facilities that may have received affected products for awareness, and complete and return the medical device response form. If further information relevant to this issue becomes necessary CONMED will communicate as such.

Distribution

Nationwide distribution

Quantity

29,403 packs (176,418 eaches)