Siemens IMMULITE 2000 XPi Immunoassay System - In vitro Chemistry Analyzer SMN #10373214 - Product Usage: are intended for professional use in a laboratory environment only.
Recall
- Recall Number
- Z-2470-2020
- Event Number
- 85731
- Firm
- Siemens Healthcare Diagnostics Inc.
- FEI Number
- 1000122204
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 15, 2020
- Terminated
- August 12, 2022
- Address
- 62 Flanders-Bartley Rd., Flanders, NJ, 07836-4715
Description
Siemens IMMULITE 2000 XPi Immunoassay System - In vitro Chemistry Analyzer SMN #10373214 - Product Usage: are intended for professional use in a laboratory environment only.
Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
Siemens issued an Urgent Medical Device Correction (UMDC IMI20-01.A.US) to US customers via FedEx on May 15, 2020. Urgent Field Safety Notice (UFSN IMI20-01.A.OUS) was electronically issued to all OUS Siemens sites with affected customers, for distribution, as well as regional reporting to their respective local competent authority on May 15, 2020. The field action letters explain the issue, health risk and and inform customers of actions to take: Siemens Customer Service Engineers shall be visiting customer sites to replace Waste Bottle Flex Tubing Assemblies. Customers are requested to return the Effectiveness Check Form, in the UMDC/UFSN via fax or email.
Worldwide distribution - US Nationwide distribution and the countries of AE, AO, AR, AT, AU, AZ, BA, BD, BE, BG, BH, BO, BR, BS, BY, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GP, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KR, KW, KZ, LB, LT, LU, LV, LY, ME, MK, MM, MQ, MT, MU, MX, MY, NL, NO, NZ, PE, PH, PK, PL, PT, RO, RS, RU, SA, SE, SG, SI, SK, SY, TH, TN, TR, TT, TW, UA, UY, VE VN, ZA.
2925 units