Spider Flex Monopolar Hook (Spider Surgical Instruments), TransEnterix, Inc., Durham, NC
Recall
- Recall Number
- Z-2465-2010
- Event Number
- 56179
- Firm
- Transenterix, INC
- FEI Number
- 3007593944
- Product Code
- GCJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 22, 2010
- Posted
- September 22, 2010
- Terminated
- January 31, 2011
- Address
- 3908 Patriot Dr, Ste 170, Durham, NC, 27703-8226
Description
Spider Flex Monopolar Hook (Spider Surgical Instruments), TransEnterix, Inc., Durham, NC
The hook end effecter from a monopolar device may become dislodged and fall into the patient when in use during laproscopic surgery. Retrieval from the patient may involve extended anesthesia time.
TransEnterix issued a "Field Action" letter dated June 22, 2010 to consignees, identifying the affected product and actions to be taken by the customer. Customers were instructed to return the product to TransEnterix, Inc. using the address label and return information provided. The firm will replace the product at no charge. TransEnterix Customer Care can be contacted at 1-888-879-4111.
Nationwide Distribution in the states of AZ, FL, GA, NC, NY, OH, OR, SC and TX.
46 boxes (6 units per box)