BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sites, REF: CM42500E-07; BD Gravity IV Set 3-port Closed Stopock Manifold Back Check Valve 2 Needle-Free Y-sites, REF: CS42522E-07; BD GRAVITY 10 DP V/NV CHECK VALVE 3-WAY STOPCOCK 4 Smartsite VALVES & SmartSite EXT SET, REF: 42407E-05
Recall
- Recall Number
- Z-2463-2021
- Event Number
- 88203
- Firm
- CAREFUSION
- FEI Number
- 3012410077
- Product Code
- FPA
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- August 2, 2021
- Address
- 10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386
Description
BD Gravity IV Set 3-port Closed Manifold Back Check Valve 3 Needle-Free Y-sites, REF: CM42500E-07; BD Gravity IV Set 3-port Closed Stopock Manifold Back Check Valve 2 Needle-Free Y-sites, REF: CS42522E-07; BD GRAVITY 10 DP V/NV CHECK VALVE 3-WAY STOPCOCK 4 Smartsite VALVES & SmartSite EXT SET, REF: 42407E-05
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
On 08/02/21, Urgent Medical Device Correction notices were mailed and emailed to consignees who were asked to notify recalling firm if this issue was experienced at 1-844-823-5433, [email protected] . Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. If during priming the clinician encounters an occluded needle-free connector, take the following actions: 1) Depress the syringe plunger while the syringe is still attached to the needle-free connector. 2) If unable to depress the plunger, pull back on the syringe plunger and depress the syringe plunger again. Repeating this step may assist in opening the needle-free connector valve. 3) If the valve still is not open; without disconnecting the syringe male luer tip from the BD SmartSite valve surface, carefully unthread and rethread the syringe to partially disengage and reengage the syringe. Repeating this step up to three times may assist in opening the needle-free valve. Note: To maintain sterility, do not allow the syringe and needle-free valve to completely disconnect. 4) If previous steps do not open the valve, then open a new needle-free connector to replace the occluded connector. 5) After the steps above, proceed with the existing instructions for use. 6) Share this correction notification with all users within your facility network of the product to ensure they are also aware of this correction. 7) Complete and return the Customer Response Form. Distributors were asked to identify customers within their distribution network that purchased the affected product and to email an excel file listing of all customers to the recalling firm within 72 hours of receipt of this letter so the recalling firm can initiate customer notification. See online tool to support the identification of impacted lot numbers https://www.bd.com/MDS-21-4144
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.
24,360