FDA Recall Terminated

GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that is intended to be mounted to a GCX compatible mount or "arm" and provide a data connection to an authorized manufacturer's patient monitoring systems.

Recall: Z-2462-2010 · Initiated August 20, 2010

Recall

Recall Number
Z-2462-2010
Event Number
56560
Firm
Abbott Point of Care Inc.
FEI Number
2245578
Product Code
CGA
Status
Terminated
Root Cause
Device Design
Initiated
August 20, 2010
Posted
September 23, 2010
Terminated
July 25, 2012
Address
400 College Rd E, Princeton, NJ, 08540-6607

Description

GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that is intended to be mounted to a GCX compatible mount or "arm" and provide a data connection to an authorized manufacturer's patient monitoring systems.

Reason

Reliability issues associated with broken connection pins within the downloader that mate with the pins from the analyzer.

Action

Abbott Point of Care, Inc. sent Urgent Recall Notices, dated August 20, 2010, by Federal Express to two Clarion Health locations. The letter identified the product, the issue, and the actions to be taken by the firm and the customers. 1) An Abbott Point of Care representative will contact the customers to plan the replacement of affected product at their facility. 2) Replacement will occur in batches of approximately 50 units at a time. 3) Abbott will provide personnel to perform the replacements to help expedite the replacement. 4) The customer is required to acknowledge that all units have been replaced by signing the form provided with the Urgent Recall Notice.. If you have any questions regarding this information please contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point of Care representative.

Distribution

Nationwide Distribution - USA, including the state of Indiana

Quantity

277