FDA Recall Terminated

Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.

Recall: Z-2461-2015 · Initiated May 1, 2015

Recall

Recall Number
Z-2461-2015
Event Number
71556
Firm
Ferno-Washington Inc
FEI Number
1523574
Product Code
FPO
Status
Terminated
Root Cause
Process control
Initiated
May 1, 2015
Posted
August 25, 2015
Terminated
November 6, 2015
Address
70 Weil Way, Wilmington, OH, 45177-9371

Description

Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.

Reason

The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.

Action

Ferno's first action of May 1, 2015 was to telephone all customers to alert them to the potential issue and explain our planned inspection. Ferno's Technical Support Manager emailed a copy of the Service Notice and iNX Equipment Evaluation Form to the customers as well.

Distribution

Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.

Quantity

26 units