FDA Recall Terminated

Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x.

Recall: Z-2461-2010 · Initiated September 30, 2008

Recall

Recall Number
Z-2461-2010
Event Number
49594
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
MOS
Status
Terminated
Root Cause
Other
Initiated
September 30, 2008
Posted
September 22, 2010
Terminated
September 30, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x.

Reason

The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination with high SAR levels and when the RF cables of the coil are too close to the patient, this RF interaction has resulted in RF burns to patients.

Action

Philips Healthcare issued URGENT-Device Correction Notices dated September 30, 2008 to their consignees. The notice identified the recalled device as 1.5 Intera and 1.5T Achieva MRI systems, described the risk to patients, and the actions to be taken by customers.. The firm has issued Field Change Order (FCO) 78100242. Field service representatives will visit each consignee and conduct a hardware upgrade, applying a sleeve over the cable that will ensure proper spacing between the cable and patient, reducing the chance of RF interaction and burns. The consignees should contact the Philips Call Center at 1-800-772-9377, #5 #3, #1 and reference "FCO 78100242 - PATIENT CABLE SLEEVE FOR 1.5T SENSE XL TORSO COIL".

Distribution

Nationwide Distribution: Throughout USA. No distribution to foreign consignees.

Quantity

47 units total (Achieva 1.5T and Intera 1.5T)