Sterile Disposable Connectors ASY 1/2 CONN W/LL STRLE 24 REF/ GTIN for Insert Label:050608000/ 00803622123382
Recall
- Recall Number
- Z-2457-2020
- Event Number
- 85623
- Firm
- LivaNova USA Inc.
- FEI Number
- 1718850
- Product Code
- DTL
- Status
- Terminated
- Root Cause
- Packaging process control
- Initiated
- April 24, 2020
- Terminated
- February 17, 2022
- Address
- 14401 W 65th Way, Arvada, CO, 80004-3503
Description
Sterile Disposable Connectors ASY 1/2 CONN W/LL STRLE 24 REF/ GTIN for Insert Label:050608000/ 00803622123382
The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.
The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.
US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)
240 units