Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

Recall: Z-2456-2015 · Initiated July 20, 2015

Recall

Recall Number
Z-2456-2015
Event Number
71815
Firm
Trilliant Surgical Ltd.
FEI Number
3007420745
Product Code
HTO
Status
Terminated
Root Cause
Component design/selection
Initiated
July 20, 2015
Posted
August 24, 2015
Terminated
January 26, 2017
Address
6721 Portwest Dr, Ste 160, Houston, TX, 77024-8019

Description

Concave Reamer Head, 16 mm; Concave Reamer Head, 18 mm; Concave Reamer Head, 20 mm; Concave Reamer Head 22 mm; Convex Reamer Head, 16 mm; Convex Reamer Head, 18 mm; Convex Reamer Head 20 mm; Convex Reamer Head 22 mm, used with Gridlock Plating system, for use in trauma and reconstructive procedures in small bones.

Reason

The cannulation of the Concave and Convex Reamer Heads may be too small to allow the Guide Pin to freely pass through.

Action

The recalling firm sent out written notification of the recall on 7/3/14. The letter instructed the consignees to exchange affected instruments.

Distribution

Nationwide throughout the US

Quantity

638 units