FDA Recall Open, Classified

Micro Joystick R-net. Electrical wheelchair component.

Recall: Z-2454-2025 · Initiated July 10, 2025

Recall

Recall Number
Z-2454-2025
Event Number
97354
Firm
mo-Vis BVBA Biebuyckstraat 15D Nevele Belgium
FEI Number
3011795161
Product Code
KNN
Status
Open, Classified
Root Cause
Device Design
Initiated
July 10, 2025
Posted
September 2, 2025

Description

Micro Joystick R-net. Electrical wheelchair component.

Reason

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Action

An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing [email protected] to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to [email protected] or by phone at +32 9 335 28 60.

Distribution

US Nationwide distribution in the state of TX.

Quantity

3 units