FDA Recall Terminated

Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

Recall: Z-2453-2012 · Initiated September 14, 2012

Recall

Recall Number
Z-2453-2012
Event Number
63180
Firm
Hologic, Inc.
FEI Number
1221300
Product Code
JAA
Status
Terminated
Root Cause
Software design
Initiated
September 14, 2012
Posted
September 25, 2012
Terminated
January 10, 2014
Address
35 Crosby Dr, Bedford, MA, 01730-1450

Description

Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.

Reason

Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180 around vertical axis as well as 180 around a horizontal axis.

Action

Hologic issued an Urgent: Medical Device recall letter dated September 14, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Halogen corrected the anomaly and they provided the customer with the revised version of the software. Customers were asked to update their systems in accordance with the instructions provided with the software included with the letter. Customers were also asked to complete the attached acknowledgement form and fax or mail it to the firm at their earliest convenience. For any questions customers were instructed to call Hologic's Help Desk at 1-800-321-4659. For questions regarding this recall call 781-999-7300.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.

Quantity

46 units