FDA Recall Terminated

SOMATOM Emotion 16. Computed tomography X-ray system

Recall: Z-2453-2008 · Initiated August 5, 2008

Recall

Recall Number
Z-2453-2008
Event Number
49139
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
JAK
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 5, 2008
Posted
September 21, 2008
Terminated
November 7, 2008
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

SOMATOM Emotion 16. Computed tomography X-ray system

Reason

Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.

Action

The firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.

Distribution

Nationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.

Quantity

53 units