FDA Recall Open, Classified

The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturers medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.

Recall: Z-2452-2025 · Initiated July 8, 2025

Recall

Recall Number
Z-2452-2025
Event Number
97371
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
LRO
Status
Open, Classified
Root Cause
Unknown/Undetermined by firm
Initiated
July 8, 2025
Posted
August 27, 2025
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

The Medline General Surgery Tray is customized to meet requirements of the hospitals and health care professionals utilizing them. The kit is provided sterile, single-use and includes a range of components dependent upon the specific procedure. These kits will be assembled using Medline and other manufacturers medical device components and drug components in some cases. Customers specify the kit contents, quantity, and placement of individual items in the kit.

Reason

Affected kits contain recalled Stryker Color Cuff Sterile Disposable Tourniquet Cuffs. Stryker Instruments initiated this recall due to a potential manufacturing defect in which the flange may detach from the bladder at the welded connection point. This detachment compromises the cuff's ability to maintain adequate air pressure, potentially resulting in tourniquet failure and increased risk of patient hemorrhaging.

Action

On July 8, 2025, the firm communicated the recall to customers who had received affected kits. Medline instructed customers to request stickers to over-label the affected convenience kits with a warning label, indicating to remove the affected component and replace it with product from supply. No kits are to be returned.

Distribution

US Nationwide distribution tin the states of CA, FL, IA, and TX.

Quantity

1145