FDA Recall
Terminated
SOMATOM Emotion 6. Computed tomography X-ray system
Recall: Z-2452-2008
·
Initiated August 5, 2008
Recall
- Recall Number
- Z-2452-2008
- Event Number
- 49139
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 5, 2008
- Posted
- September 21, 2008
- Terminated
- November 7, 2008
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
SOMATOM Emotion 6. Computed tomography X-ray system
Reason
Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.
Action
The firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.
Distribution
Nationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.
Quantity
53 units