FDA Recall Terminated

Flexible Intubation Fiberscope, Model # 11301AB1

Recall: Z-2447-2019 · Initiated April 24, 2019

Recall

Recall Number
Z-2447-2019
Event Number
83419
Firm
Karl Storz Endoscopy
FEI Number
3007475226
Product Code
EOQ
Status
Terminated
Root Cause
Process control
Initiated
April 24, 2019
Terminated
December 4, 2020
Address
2151 E Grand Ave, El Segundo, CA, 90245-5017

Description

Flexible Intubation Fiberscope, Model # 11301AB1

Reason

The action is being initiated due to intubation scopes improperly manufactured with a double port luer instead of a single port luer.

Action

The firm mailed a "safety alert" on April 24, 2019 via FedEx. 2nd day delivery. The safety alert advised customers to do the following: 1. If you use the following automated processes, the high level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed: Reprocessing Method Impact, Medivators AER * Patient Safety may be Compromised, STERIS System 1E (SS1E)** Patient Safety may be Compromised The above connectors can only connect onto one port, therefore leaving the other port open. Depending on the reprocessing system, the cycle could result in an error or in successful completion of the cycle. " If an error code is displayed, the user would know the cycle is not complete and the scope has not been successfully reprocessed. " If the cycle is completed without an error code, the high-level disinfection/liquid chemical sterilization of the flexible endoscope cannot be guaranteed. Improper reprocessing may result in incomplete decontamination, which may result in a cross-contamination patient infection. If you use Medivators AER or STERIS SS1E to reprocess this scope, please be aware of the possible risks and take the necessary actions to address any patient safety concerns. 2. Please complete the attached Action Requested form and follow the instructions outlined on the form. 3. you have any questions please contact the following person: This Recall Action, Emily Cuadros, Complaint Handling , 424-218-8289 Reprocessing, Shaun McGinley, Director of Reprocessing, 424-218-8247 Patient Safety, Dr. Dennis Fowler, Chief Medical Officer, 424-218-8603

Distribution

US: VA, MO, CA, OH, NM, NY, CO, LA, ME, OUS: Canada

Quantity

14 scopes