Oscor PACE 203H, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.
Recall
- Recall Number
- Z-2443-2010
- Event Number
- 56121
- Firm
- Osypka Medical Inc
- FEI Number
- 3004185614
- Product Code
- DTE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- June 30, 2010
- Posted
- September 21, 2010
- Terminated
- August 10, 2012
- Address
- 7855 Ivanhoe Ave, Ste 226, La Jolla, CA, 92037-4561
Description
Oscor PACE 203H, Dual-Chamber External Pulse Generator The PACE 203H / Model 3085 external pulse generator / temporary cardiac pacemaker is designed to be used with cardiac pacing lead systems for temporary atrial or ventricular or A-V sequential pacing. The PACE 203H / Model 3085 has applications where such pacing modes are indicated for therapeutic, prophylactic, or diagnostic purposes.
The recall was initiated after Osypka Medical received two (2) complaints from their Japanese distributor regarding an incompatibility of the PACE 203H with some new types of 9 V batteries. The investigation revealed that PACE 203H of a certain manufacturing period malfunction (i.e., not stimulate) when a type of 9 V batteries with higher than usual open circuit voltage (OCV) is used. Failure t
OSYPKA Medical sent "Urgent Medical Device Recall" letters dated June 30, 2010 and July 7, 2010 to the direct customers. The recall notice informed the customers of the Reason for Recall, the Potential Hazard, the Temporary Fix of the Problem, the Permanent Fix of the Problem and Recommended Actions to be taken by the customer. The customers were instructed to contact Osypka Medical via telephone (858) 454-0021 or e-mail to [email protected] to verify that the Model 3085 and/or PACE 203H is subject to this product removal, obtain a return-to-manufacturer-authorization (RMA) number ,and then return the affected devices to Osypka Medical. If you have any questions call (858) 454-0021 or e-mail to [email protected]
Nationwide Distribution: USA including states of CA & FL
205 units (US Only)