Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Recall
- Recall Number
- Z-2441-2018
- Event Number
- 80418
- Firm
- Zeiss, Carl Inc
- FEI Number
- 2431026
- Product Code
- RGW
- Status
- Terminated
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- May 29, 2018
- Terminated
- July 26, 2024
- Address
- 1 Zeiss Dr, Thornwood, NY, 10594-1939
Description
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
The Company will immediately notify by direct contact (when available) and/or express courier each first line purchaser and/or known user site with informing them of the risk, how to prevent it and the notice of action. The Company will dispatch factory trained service representatives to perform following actions: o Installation of Hot Fix # 14 (or newer) for ZEN 2.3 o Installation of Firmware 1 .209 (or newer) on nodes x l 5 and x l 9 (Main Controller 2)
US Nationwide Distribution
8