FDA Recall Terminated

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Recall: Z-2441-2018 · Initiated May 29, 2018

Recall

Recall Number
Z-2441-2018
Event Number
80418
Firm
Zeiss, Carl Inc
FEI Number
2431026
Product Code
RGW
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
May 29, 2018
Terminated
July 26, 2024
Address
1 Zeiss Dr, Thornwood, NY, 10594-1939

Description

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Reason

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

Action

The Company will immediately notify by direct contact (when available) and/or express courier each first line purchaser and/or known user site with informing them of the risk, how to prevent it and the notice of action. The Company will dispatch factory trained service representatives to perform following actions: o Installation of Hot Fix # 14 (or newer) for ZEN 2.3 o Installation of Firmware 1 .209 (or newer) on nodes x l 5 and x l 9 (Main Controller 2)

Distribution

US Nationwide Distribution

Quantity

8