FDA Recall Terminated

Dimension Vista MULTI 1 SDIL, KD693, SMN# 10469971

Recall: Z-2437-2018 · Initiated March 21, 2018

Recall

Recall Number
Z-2437-2018
Event Number
79784
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
CHP
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
March 21, 2018
Terminated
June 12, 2020
Address
500 Gbc Dr Po Box 6101, Ms 514, Newark, DE, 19702-2466

Description

Dimension Vista MULTI 1 SDIL, KD693, SMN# 10469971

Reason

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Action

Notification letters were distributed 3/21/18. The letters instructed customers to: Inspect stock, discontinue use of and discard the Dimension Vista Diluent lots listed in Table 1. Please review this letter with your Medical Director. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens Technical Support Representative. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product.

Distribution

The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to Guam, Saipan, and Virgin Islands. The products were distributed to the following countries: Australia, Austria, Bahamas, Belgium, Brazil, Canada, Canary Islands, Croatia, Czech Republic, Denmark, France, Germany, Italy, Japan, Lebanon, Netherlands, Poland, Portugal, Qatar, Russian Federation, Slovakia, Slovenia, Saudi Arabia, South Korea, Spain, Switzerland, and United Arab Emirates.

Quantity

197