FDA Recall Terminated

Infinion is Nuclear Magnetic Resonance Imaging. Philips Infinion 1.5T System, Picker International, Inc. World Headquarters 595 Miner Road Cleveland, Ohio 44143, U.S.A..

Recall: Z-2433-2010 · Initiated September 19, 2008

Recall

Recall Number
Z-2433-2010
Event Number
49489
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
September 19, 2008
Posted
September 17, 2010
Terminated
September 17, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Infinion is Nuclear Magnetic Resonance Imaging. Philips Infinion 1.5T System, Picker International, Inc. World Headquarters 595 Miner Road Cleveland, Ohio 44143, U.S.A..

Reason

Water may collect and freeze in the vent pipe. This can rupture the magnet vessel during a cryogen transfer, and release helium into the MRI suite.

Action

Philips issued an Urgent Correction letter dated 09/19/2008 to their consignees, identifying the reason for the recall and the affected products. The firm issued the Field Change Order (FCO) 78100232. Field service engineers will visit each consignee and conduct an inspection of the magnets venting system of Edge, Vista, Eclipse, Polaris, Infinion, and Panorama 0.6T MRI systems. The consignees should contact the Philips Call Center at 1-800-772-9377, #5 #2 and reference "FCO 78100232".

Distribution

Nationwide Distribution: USA., including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY. There is no distribution to foreign consignees.

Quantity

87 units